12.3 Operating mode
Operating mode
Operation with non-periodic load and
speed changes (type S9 pursuant to
IEC EN 60034-1)
Drilling (clockwise/counterclockwise rota-
tion):
60 second application, 60 second pause
6 repetitions
30 min cooling time
Max. Temperature 48 °C
Medullary reaming (clockwise/counter-
clockwise rotation):
30 second application, 30 second pause
8 repetitions
30 min cooling time
Max. Temperature 48 °C
Drilling (oscillation):
15 second application, 15 second pause
3 repetitions
30 min cooling time
Max. Temperature 48 °C
Saw mode with GB891R:
30 second application, 60 second pause
4 repetitions
30 min cooling time
Max. Temperature 48 °C
Saw mode with GB892R:
30 second application, 60 second pause
5 repetitions
30 min cooling time
Max. Temperature 48 °C
12.4 Environmental conditions
Operation
Temperature
10 °C to 27 °C
Relative humidity
30 % to 75 %
Atmospheric pressure
700 hPa to
1 060 hPa
13. Disposal
Note
The user institution is obliged to process the product before its disposal, see
Validated reprocessing procedure.
Adhere to national regulations when disposing of or recy-
cling the product, its components and its packaging!
The recycling pass can be downloaded from the Extranet
as a PDF document under the respective article number.
(The recycling pass includes disassembling instructions
for the product, as well as information for proper disposal
of components harmful to the environment.)
Products carrying this symbol are subject to separate col-
lection of electrical and electronic devices. Within the
European Union, disposal is taken care of by the manu-
facturer as a free-of-charge service.
Detailed information concerning the disposal of the product is avail-
able through your national B. Braun/Aesculap agency, see Technical
Service.
14. Distributor in the US/Contact in Can-
ada for product information and com-
plaints
Aesculap Inc.
3773 Corporate Parkway
Center Valley, PA, 18034,
USA
Storage and trans-
port
-10 °C to 50 °C
10 % to 90 %
500 hPa to 1 060 hPa
19