9.
Troubleshooting list
Have defective products repaired by Aesculap Technical Service, see Technical Service.
Issue
Cause
Product becomes
Excessive use
too hot
Defective attachment
drive/ball bearings
Fall damage, product
defective
Blunt tool
Insufficient
Product is operating in
power
left rotation mode
Attachment defective
Blunt tool
Loud running
Defective attachment
noise
drive/ball bearings
Attachment does
Attachment not compati-
not attach or
ble
detach.
Connection on attach-
ment deformed/defective
Coupling on drill
deformed/defective
Sleeve on drill hard to
move
Tool does not
Incompatible tool
attach or detach.
Tool connector
deformed/defective
Coupling on attachment
deformed/defective
Tool coupling dirty
Detection
Attachment very hot
Attachment very hot
Attachment very hot
Heating of tool and product
Serrated tool run in counterclockwise
mode
Attachment very hot
Tool blades worn out
Loud, perceptible noise during operation
Attachment doesn't engage
Attachment is difficult to couple or can-
not be coupled
Attachment is difficult to couple or can-
not be coupled
Attachment doesn't engage
Tool does not click into place
Tool is difficult to couple or cannot be
coupled
Tool is difficult to couple or cannot be
coupled
Tool is difficult to couple or cannot be
coupled
Remedy
Follow operating instructions (operating mode).
Follow operating instructions (preparation, care).
Preventative: lubricate attachment before every steril-
ization.
Have the manufacturer repair the attachment.
Have the attachment repaired by the manufacturer.
Change tool.
Use serrated tools only in clockwise mode.
Follow operating instructions (preparation, care).
Preventative: lubricate attachment before every steril-
ization.
Follow operating instructions (operating mode).
Have the attachment repaired by the manufacturer.
Change tool.
Follow operating instructions (preparation, care).
Preventative: lubricate attachment before every steril-
ization.
Have the manufacturer repair the attachment.
Use the compatible attachment for drill.
Have the manufacturer repair the attachment.
Have the manufacturer repair the drill.
Turn and hold the drill sleeve, then attach the attach-
ment. Turn the sleeve back if necessary.
Preventative: lubricate the sleeve on the drill before
each sterilization.
Use the tool compatible with the attachment.
Use a new tool.
Have the manufacturer repair the attachment.
Clean tool or use a new tool.
Clean the attachment.
15