Uncoupling
Turn sleeve 4 in the direction of the arrow 5 and simultaneously
remove the attachment 6/9/12/16/18/24 from the drill 1.
6.1.4
Attaching and detaching the tool from the attachment
Risk of injury when attaching/removing attach-
ments/tools in the ON position through inadvertent
activation of the product!
WARNING
Only attach/remove attachments/tools in the
OFF position.
Ensure that the tool connector and attachment type are compatible.
Attachments with quick-mount chuck GB871R/GB872R/
GB884R/GB886R/GB887R
Coupling:
Pull back on the unlocking sleeve 7.
Insert the tool shaft in the correct position up to the stop in the tool
holder 8 of the attachment 6.
Release the unlocking sleeve 7.
Tool is attached.
Pull at tool to check that it is securely coupled.
Uncoupling:
Pull back the unlocking sleeve 7.
Remove tool.
Attachments with three-jaw chuck GB870R/GB881R/GB882R
Coupling:
Open the three jaw chuck with the wrench 10.
Insert the tool shaft in the correct position up to the stop in the tool
holder 11 of the attachment 9.
Close the three jaw chuck 10 and pull tight.
Pull at tool to check that it is securely coupled.
Uncoupling:
Open the three jaw chuck with the wrench 10.
Remove tool.
Kirschner wire attachment GB894R
Risk of injury when using long guide wires!
Use Kirschner wire protection sleeve when using
long guide wires.
WARNING
Note
The special Kirschner wire chuck is recommended for the placement of
guide wires. This quick-action chuck adapter allows quick and easy ten-
sioning of drilling wires.
When using long guide wires, a Kirschner wire protection sleeve must be
used with the drill to prevent injury.
For further information on GA344, see TA014550 and TA014551 (leaflet).
For further information on GA844, see TA014436 and TA014437 (leaflet).
The following diameters can be set on the Kirschner wire attachment:
0.6 mm to 1.2 mm
1.2 mm to 2.2 mm
2.2 mm to 3.2 mm
Using a Kirschner wire:
Ensure that the clamping lever 13 is in the starting position (deacti-
vated status).
Set the adjusting sleeve 14 of the Kirschner wire attachment 12 to the
desired diameter range:
– Push back the adjusting sleeve 14 and turn until the desired diam-
eter range has been reached.
– Release the adjusting sleeve 14. Ensure that the adjustment
sleeve 14 clicks into place.
Insert the Kirschner wire into the Kirschner wire chuck 15 until the
desired exposed length is achieved.
The Kirschner wire is held in the intended position in the attachment
by an automatic clamping mechanism.
Tension the Kirschner wire:
Pull on the clamping lever 13 and hold in the desired position.
The further the clamping lever is pulled back, the higher the tension of
the Kirschner wire.
Note
The Kirschner wire will only remain tensioned with the clamping lever
pulled back. If the clamping lever is released, it will move back into the
starting position and the Kirschner wire is freely mobile.
Drill attachment for radiolucent angle transmission GB878R
The drill attachment allows connecting a radiolucent angle transmission
for applying drill holes in the bone under radiographic control.
The drill attachment is only suitable for use of the radiolucent angle trans-
mission 511.300 from the Synthes company.
Follow the instructions for the radiolucent angle transmission 511.300
in the operating instructions of the Synthes company.