Attachments for small drill GA344 and drill GA844
Issue
Cause
For GB894R:
Kirschner wire chuck not
Kirschner wire
set correctly
does not insert
into the Kirschner
wire chuck
For GB894R:
Kirschner wire chuck not
Kirschner wire
set correctly
not turning
Open the clamping lever
Clamping lever not pulled
back with sufficient force
Tool not moving
Attachment not fully
attached to the drill
Tool not fully attached to
the attachment
Attachment defective
Drill defective
The safety catch of the
drill is in the OFF position
Oil spray adapter
Oil spray adapter not
not attachable
compatible
16
Acculan 4 / ELAN 4 electro
®
Detection
Diameter of the Kirschner wire is not
consistent with the scale setting
Diameter of the Kirschner wire is not
consistent with the scale setting
Open the clamping lever
Kirschner wire not turning
Attachment can be pulled out of the drill Correctly attach the attachment and perform a function
Tool can be pulled out of the tool cou-
pling
Drill turning, but attachment is not turn-
ing.
Drill not turning
Safety catch is in OFF position
Oil spray adapter not attachable
Remedy
Set the adjustment sleeve to the right Kirschner wire
diameter.
Set the adjustment sleeve to the right Kirschner wire
diameter.
Close the clamping lever.
Pull back the clamping lever forcefully and hold it.
test.
Correctly attach the tool and perform a function test.
Have the manufacturer repair the attachment.
Have the manufacturer repair the drill.
Turn the safety catch into the ON position.
Use the compatible oil spray adapter for GA344.