12.1
General faults
Fault
Required function not avail-
able
(
)
seca 496
Display does not react when
keys are pressed
Display remains dark after
pressing or switching on
12.2
Measuring weight (seca 495/495 r)
Fault/error code
Displayed weight is implausi-
ble
to
E02.01
E02.08
112 • seca 604, seca 495/495 r/496, seca 404/405, seca 446/448, seca 442, seca 438, seca 476
Information about display messages and the structure of error codes can be
found here:
è Traffic light system for display messages (seca 496), page 119
è Error codes (basic display), page 119
è Error codes (ID display), page 119
Cause
Device mode in which the function is
not provided is active
Device is in an undefined state follow-
ing implausible input
No connection to weighing platform
No power supply:
• No plug-in power supply unit con-
nected (
seca 495/495 r/496
• No batteries inserted (
seca 495/495 r
Device is in an undefined state follow-
ing implausible input
Display faulty
Cause
Weighing electronics using outdated
zero point
Weighing electronics defective
Load cell or weight calculation module
defective
Clarify with the administrator/hospital tech-
nician whether the device can be operated
in a different device mode
è Changing device mode, page 90
•
•
è Functions/Device mode (seca
496), page 124
• Restarting the device
• If the error recurs, inform seca Service
• Check whether the display cable is con-
nected correctly
• If the error recurs, inform seca Service
• Check whether the plug-in power supply
unit is connected correctly
(
seca 495/495 r/496
)
• Check whether the batteries are inserted
correctly (
seca 495/495 r
)
• If the error recurs, inform seca Service
• Restarting the device
• If the error recurs, inform seca Service
Inform seca Service
• Remove the weight from the weighing
platform
• Restarting the device
• Wait until main screen is displayed again
Inform seca Service
Inform seca Service
Remedy
)
)