Descargar Imprimir esta página

Seca 604 Instrucciones De Uso página 90

Publicidad

Idiomas disponibles

Idiomas disponibles

Stand-alone operation
9
CONFIGURATION (ID DISPLAY)
9.1
Basic functions
Changing device mode
90 • seca 604, seca 495/495 r/496, seca 404/405, seca 446/448, seca 442, seca 438, seca 476
1. Press the
key.
ð The measuring results are submitted to the EMR system and assigned to
the electronic patient file.
2. Ask the patient to step off the weighing platform.
ð The device is ready for the next measurement.
To complete a measurement procedure on devices that are not connected to an
EMR system, proceed as follows:
1. Make sure that the displayed measured values are plausible.
2. Transfer the displayed measured values to the patient file manually.
3. Ask the patient to step off the weighing platform.
4. Press the
key.
ð Measured values and manual entries are discarded.
ð The device is ready for the next measurement.
NOTE
– The functions described in this part of the instructions for use are
intended exclusively for administrators and hospital technicians.
– This part of the user documentation contains information about
configuring the device for measuring mode and for integrating it into
a PC network.
NOTICE!
Integrating this device into a PC network containing other devices may
lead to previously unknown risks for patients, operators or third par-
ties.
► It is the responsibility of the operating company to determine, ana-
lyze, rate, and manage these risks.
The following device modes are available for setting the device up for different
usage situations:

Publicidad

loading

Este manual también es adecuado para:

495496404405446448 ... Mostrar todo