Message
icon appears
icon appears
key appears grayed out
key appears grayed out
Autoclear function: An entry
of "0 seconds" is not ac-
cepted; the factory setting
(300 seconds) is suggested
instead
Workflow LED does not light
up during a measurement or
while data are to be transmit-
ted
Workflow LED illuminated red
001-318XX-XXX
002-318XX-XXX
004-318XX-XXX
Cause
Optional measured value (e.g. height)
has not been determined
Optional ID (e.g. user ID) has been
scanned
There are no data available to be con-
firmed
There are no data available to be
cleared
Implausible input; switching off the
function is not intended in the factory
No network connection
WiFi function deactivated
Workflow LED defective
ID was
found in the EMR system or
not
the seca software
Device did
save the measuring re-
not
sults to the clipboard
Measuring results were
not
EMR system or a seca software
No connection to the server
No data transmission possible
Barcode scanned at the wrong time
• Press
measured value
• Press
• If desired: Check workflow settings for the
device in the
If desired: Check workflow settings for the
device in the
• Perform a measurement
• Scan IDs (user/patient)
• Perform a measurement
• Scan IDs (user/patient)
• Accept factory setting
• Manually enter value between 1 and 3600
seconds
Set up network connection
network
functions, page 101)
Activate the WiFi function
deactivating the WiFi
Inform seca Service
Create the ID in the EMR system or in the
seca software
Repeat measurement
• Repeat measurement
sent to an
• Check WiFi connection
• Check LAN cable
• Check network settings
network
Check network and workflow settings:
• Check network settings
network
• Check workflow settings: System instruc-
tions for use
Repeat measurement and maintain correct
operating sequence
Restarting the device
If the error recurs, inform seca Service
English
Remedy
key and determine optional
key and end measurement
software
seca connect 103
software
seca connect 103
(è Setting up
(è Activating/
function, page 103)
(è Setting up
functions, page 101)
(è Setting up
functions, page 101)
seca 103/452
Troubleshooting • 115