Descargar Imprimir esta página

Seca 604 Instrucciones De Uso página 48

Publicidad

Idiomas disponibles

Idiomas disponibles

5.17
Mounting the scanner bracket to the handrail
5.18
Setting up and aligning the device
48 • seca 604, seca 495/495 r/496, seca 404/405, seca 446/448, seca 442, seca 438, seca 476
WARNING!
Injury from falling
► Position the scale at a suitable location with little passing traffic.
Mount the ramps supplied on the intended sides (except next to
the wall) so no-one can trip over them.
WARNING!
Injury from falling
The patient handling aid might get caught in gaps and the patient
could fall out of the patient handling aid.
► Make sure that the area for setup is clean and level.
► When setting up the device, follow the installation instructions and
ensure the smallest possible offsets and gaps between the weigh-
ing platform and the ramps and between the ramps and the floor.
WARNING!
Faulty measurement
► Keep a sufficient distance from walls or objects to avoid faulty mea-
surements. The weighing platform, the guardrail and the handrail (if
present) should not touch anything else. Do not press the scale
against the wall.
WARNING!
Faulty measurement
Weighing platform bends under loading
► Place the weighing platform on level ground. Under loading, the
weighing platform must maintain a sufficient distance from the
ground.
► Make sure that no elevations are larger than 4.5 mm at the lowest
foot setting on the ground.
Lowest
foot setting
► Make sure that only the feet are in contact with the ground.
Lowest
foot setting
max. 4.5 mm

Publicidad

loading

Este manual también es adecuado para:

495496404405446448 ... Mostrar todo