Setting filtering (FIL)
74 • seca 604, seca 495/495 r/496, seca 404/405, seca 446/448, seca 442, seca 438, seca 476
1. Press the
 key for approx. 4 seconds.
ð The menu is displayed.
2. Select the
CLEAr
menu item.
3. Confirm your selection.
ð The current setting is displayed.
4. Select the desired setting:
: The display value clears after the set time expires
3
Use the arrow keys to set a time between 30 s and 1800 s.
2
: A new load on the weighing platform clears the display value
: Removing the load from the weighing platform clears the display
1
value
1. Confirm your selection.
ð The setting is adopted.
ð The menu closes.
Setting the filter (
FIL
= filter) allows you to reduce interference when determining
weight. The selected setting affects the sensitivity with which the weight display
reacts to patient movements and the time until the
weight value permanently(
Setting (
)
Sensitivity
FIL
0
Sensitive
Medium
1
2
Insensitive
NOTE
With the
0
setting and patients who are not very steady on their feet, it
is possible that no weight value is displayed permanently despite the
Hold
function being activated.
The greatest deviation between weight value displayed and actual
value occurs at setting
1. From the menu, select the
1. Confirm your selection.
ð The current setting is displayed.
1. Select a filtering level.
1. Confirm your selection.
ð The setting is adopted.
ð The menu closes.
Hold
value).
Hold
Determination of
value
Slow
Medium
Fast
.
2
menu item.
FIL
function displays a
Hold