Chapter 2
48
User Interface
English 8930 System User Guide 2010-12
MA07445A008
Rev X
Caution Messages
Table 2-8. Caution Messages
No.
Message
51
Urethra temperature high
Irrigate
52
Left needle impedance (Ω)
low
53
Right needle impedance
(Ω) low
Medtronic Confidential
CRMRef_R06
Explanation
Urethra temperature
■
°
exceeds 43
C and must
be irrigated with cool water
to lower temperature
Message stays on until
■
temperature drops below
°
43
C
Alarm is 3 short beeps
■
repeated every 4
seconds
If temperature reaches
■
°
47
C, system stops
lesion (see alert message
34, Table 2-7 on page 42)
Occurs only at start of a
■
lesion when left needle has
impedance < 80 ohms
Nominal starting range is
■
80 to 180 ohms
Physician may continue
■
with lesion
Message stays on for ten
■
seconds
Alarm is 2 short beeps
■
Occurs only at start of a
■
lesion when right needle
has impedance < 80 ohms
Nominal starting range is
■
80 to 180 ohms
Physician may continue
■
with lesion
Message stays on for ten
■
seconds
Alarm is 2 short beeps
■