Getting Started
Internal System Error Shutdown
English 8930 System User Guide 2010-12
MA07445A008
Rev X
If the system's internal monitoring detects a system error, it
automatically halts the Prostiva RF Therapy application. When
this occurs, the LCD display screen will also clear and display a
solid blue image over the entire screen. If radio frequency energy
is on, it automatically shuts off.
An internal system error may be caused by a temporary
interference. It could also be due to a permanent hardware
problem.
Complete the following steps if an internal system error shutdown
occurs:
1. Stop the treatment session; if doing a lesion, retract the
needles, and withdraw the hand piece.
2. Record the error code or message that displays on the
screen; this is for the Medtronic service representative.
3. Turn off the AC power; wait approximately five seconds and
then turn on the power.
4. If the solid blue image and error message persists on the
screen, turn off the AC power and contact your Medtronic
service representative.
Refer to "Shutting Down System, Disconnecting Cables and
Tubing System" on page 81 for instructions on how to disconnect
cables and tubing system components.