• Failure to use the Charger with either
the Charging Belt or an adhesive
patch, as shown, may result in a
burn. If pain or discomfort is felt,
cease charging and contact Boston
Scientific.
The Charger Base Station should be plugged
in and the Charger placed in the Base Station
when not in use. When the indicator light is
green, the Charger is fully charged. When
the indicator is amber, the Charger is partially
charged, but is still able to deliver a charge to
the Stimulator.
1. When the indicator light is green, remove
the Charger from the Base Station. The
indicator light will then turn off.
2. Press the power button. The indicator
light will come on again, and the Charger
will begin beeping as it searches for the
Stimulator.
3. Place the Charger over the Stimulator.
When the Charger is aligned with the
Stimulator, the beeping will stop.
• Centering the Charger over the
Stimulator will ensure the shortest
charging time.
• Many patients are able to feel the
implanted Stimulator and can place
the Charger directly on top of it.
• Alternatively, centering the Charger
within the alignment area (i.e.,
the area where the Charger does
not beep) will also ensure that the
Charger is centered.
Rechargeable Stimulator System
4. Secure the Charger over the Stimulator
by using either an adhesive patch or the
Charging Belt.
• Adhesive Patch: Remove the clear
liner from the patch. Apply the white
side with the blue stripe to the back
of the Charger. Then remove the
beige liner from the patch. Secure
the Charger over the Stimulator by
pressing the adhesive to the skin over
the Stimulator (see diagram below).
• Charging Belt: Place the Charger into
the pocket on the Charging Belt with
the power button facing out. Secure
the Charger over the Stimulator by
adjusting the Charging Belt (see
diagram below).
9055959-008 Rev A 39 of 327
Clinician Manual