The Remote Control will display the contact
impedance status screen.
Contacts 1 through 8 (lead position "1-L") are
represented by the rectangles in the top row,
starting with contact 1 on the left. Contacts
9 through 16 (lead position "2-R") are
represented by the rectangles in the bottom
row, starting with contact 9 on the left.
Rechargeable Stimulator System
Contacts within the typical impedance range
are displayed as solid rectangles. High
impedance contacts (above 4500 ohms) are
represented by hollow rectangles.
To display the impedance values press any
key (other than the Stim On/Off button
Impedance values for contacts 1 - 8 will be
shown first. Press the key again to show
impedance values for contacts 9 - 16.
9055959-008 Rev A 49 of 327
).
!
Clinician Manual