Lead Extension
Lead Extensions are designed to connect
the percutaneous and paddle leads to the
Precision IPG for spinal cord stimulation.
The extension may be added to a lead to
externalize the lead for a trial procedure or
to extend the lead when a permanent IPG is
implanted.
Lead extensions are available in lengths of
25 cm, 35 cm, and 55 cm. Each extension
has eight electrode contacts located near
the distal end. Each contact is 3 mm in
length and is spaced 1 mm from the adjacent
contact. The extension can be connected to
either the Trial Stimulator (via an OR Cable
Assembly included in the Leads Kit) or
directly to the Precision IPG.
Lead Splitter
The 2x4 Splitters are designed to connect
multiple percutaneous leads to the Precision
IPG. The Linear leads may be inserted
into a splitter for a maximum of four Linear
leads per IPG. Four of the eight contacts
of each Linear lead will be activated. Two
configurations of 2x4 splitters are available:
Distal 4 (D4) and Wide 4 (W4). The two
versions offer different contact configurations.
The Lead Splitter 2x8 is required to connect
the Infinion™ 16 contact lead to the IPG. Only
one Infinion™ 16 lead through one Splitter
2x8 can be connected to a Precision IPG.
Indications for Use
Indications for Use
The Precision™ System is indicated as an
aid in the management of chronic intractable
pain.
Contraindications
Patients contraindicated for permanent SCS
or PNS therapy are those who:
• are unable to operate the SCS or PNS
system
• have failed trial stimulation by failing
to receive effective pain relief
• are poor surgical risks
• are pregnant
9055959-008 Rev A 7 of 327