Notice - Potential damage to parts!
Closing the ball valve before you remove the flowmeter sensor can seriously damage the
protective cage or the sensor elements.
Do not close the ball valve until the flowmeter sensor has been removed.
Fig. 8:
Dimensions in mm (inch)
1
Centering pin
2
Sealing ring groove
Flowmeter sensor length h
in mm (inch)
263 (10.35)
425 (16.73)
775 (30.51)
1)
This maximum pipe diameter specification is only valid when installing the sensor unit centrically in the pipe.
For larger diameters or angular ducts a non-centric sensor position is taken into account for calibration.
Important (Notice)
Deviations from the stated dimension and position tolerances cause additional measuring
uncertainty.
CI/FMT500-IG-X1
2
1
Ø d
min. 28 (1,10)
← Direction of flow
Outer pipe diameter min. / max.
in mm (inch)
100 ... 150 (3.94 ... 5.91)
> 150 ... 500 (5.91 ... 19.69)
> 500 ... 1150 (19.69 ... 45.28)
Sensyflow FMT500-IG
1
Ø 48,3 (1,90)
3
Connection flange DN 25 (1")
D
Outer pipe diameter
1)
Mounting
3
G00803
EN - 17