DEVICE MALFUNCTION OR FAILURE
If any component of the Cellfina
immediately. Do not attempt to repair device.
STORAGE, TRANSPORT AND HANDLING ENVIRONMENTAL CONDITIONS
Store Cellfina System at room temperature, in a clean, dry location. Avoid prolonged exposure to elevated
temperature.
The Temperature, Humidity and Atmospheric Pressure conditions listed below are recommended.
Ambient temperature range of - 18° C to +60° C;
Relative humidity range of 15 % to 90 %, including condensation.
Atmospheric pressure range of 70 kPa to 106 kPa.
USE CONDITIONS
Use Cellfina System at room temperature.
LIST OF APPLIED PARTS
Reversible Lid
Release Guidance Platform
Vacuum Handpiece Body
SPECIFICATIONS
Power Supply Voltage - Input ...................................................................................... 100 ~ 240 VAC
Power Supply Voltage - Output per Connector ........................................................... 5 VDC
Power Supply Current - Output ................................................................................... 3.0A
Power Supply Power - Output ..................................................................................... 15.0W
Lead Free Status ......................................................................................................... Lead Free
Restriction of Hazardous Substances (RoHS) Status .................................................. RoHS Compliant
Motor Voltage ............................................................................................................... 3.6 VDC
Motor Reciprocations Per Minute ................................................................................. 3000
Power Supply Model Number and Manufacturer .......................................................... Globetek: GTM41060-1505
Power Supply Alternate Model Number and Manufacturer ......................................... Mean Well GEM18I05-P1J
Power Supply Class ..................................................................................................... Class II
Applied Part Type ......................................................................................................... BF
Motor Module Duty Cycle ............................................................................................. 30s On/10s Off
Ingress Rating .............................................................................................................. IP21
1007270IFU Rev D
®
Cellfina
System
®
System is associated with a malfunction or failure, please contact Ulthera
Blade
Teardrop Guidance Platform
Narrow Guidance Platform
User Manual
25 of 214