Installation in the vicinity of pumps

For flowmeter primaries which are to be installed in the vicinity of
pumps or other vibration generating equipment, the utilization of
mechanical snubbers is advantageous.
Fig. 16
4.5.8

Installing the high temperature design

The high temperature design allows for complete thermal insulation
of the sensor. The pipeline and sensor must be insulated after
installing the unit according to the following illustration.
Fig. 17
1
Insulation
4.5.9

Devices with extended diagnostic functions

For devices with extended diagnostic functions different installation
conditions may be valid.
For further information read and observe chapter 8 "Extended
diagnostic functions".
Change from one to two columns
CI/FEX300/FEX500-X1
G00561
1
G00654
FEX300, FEX500
4.5.10 Installation in pipelines with larger nominal
diameters
Determine the resulting pressure loss when using reduction pieces
(1):
1. Calculate the diameter ratio d/D.
2. Determine the flow velocity based on the flow range nomograph
(Fig. 19).
3. Read the pressure drop on the Y-axis in Fig. 19.
Fig. 18
1
= Flange transition piece
d
= Inside diameter of the flowmeter
V
= flow velocity [m/s]
Δp
= pressure loss [mbar]
= Inside diameter of the pipeline
D
Nomograph for pressure drop calculations
For flange transition piece with α/2 = 8°
Fig. 19
Mounting
EN - 21