LED light source OP950 with sterile adapter OP941
Risk of injury and/or product malfunction due to
incorrect operation of the electromedical system!
►
WARNING
Adhere to the instructions for use of any medi-
cal device.
►
Do not use the product in a magnetic resonance
environment.
4.1
Set-up environment/set-up location
Fire and explosion hazard!
►
Use the product outside of explosion hazard
zones (e.g. areas with highly pure oxygen or
DANGER
anesthetic gases).
►
Make certain that the power plug is connected
to the power supply outside of any areas where
there is a risk of explosion.
►
Have ready replacement products/light source.
The device is authorized for operation in the operating room.
Note
After being installed and put into operation, the device must not be trans-
ported or moved to a different set-up location.
Note
If the equipment cart is set up outside of the sterile area, measure the cable
length to give enough room for movement so that the sterility is not com-
promised.
►
Make certain that the ventilation slots in the bottom of the housing
and in the rear panel of the device are not covered, e.g. by an OR cloth.
►
Make certain that the following minimum distances from other devices
are maintained:
– on the sides 50 mm
– from the rear 60 mm
►
Make certain that the control elements, mains power switch and
power socket are freely accessible for the user.
►
Make certain the system is set up on a sufficiently stable support (e.g.
a table, ceiling support, equipment cart, etc.).
►
Follow the instructions for use of the support.
4.2
Stacking of units
►
Do not exceed a stack height of 415 mm.
►
Place the units in a stable position.
►
Aesculap devices must be stacked on top of each other facing identi-
cally.
►
Never stack in an offset arrangement.
6
®
5.
Working with the LED light source
5.1
System set-up
Risk of burns from excessively high operating tem-
peratures!
WARNING
The optical cable socket and optical cable connec-
tions get hot.
►
Do not situate optical cables near any flamma-
ble objects (e.g. drape).
►
Select the lowest lighting level necessary to
provide satisfactory illumination of the target
area.
Danger of blinding from light rays!
►
Never look into the open end of an optical cable
or endoscope that is connected.
CAUTION
Risk of injury if the light source fails!
►
Keep a ready-to-operate spare light source
ready.
CAUTION
5.1.1
Connecting devices for communication
The communication interfaces 8 connect the light source to a superordi-
nate device. Through these, the light source receives control signals from
the connected device.
The connection of a superordinate device to the communication interfaces
leads to the construction of a system.
Risk of injury due to unapproved configuration
using additional components!
►
Make sure that accessory equipment and/or
WARNING
peripheral devices that are connected to the
device interfaces verifiably satisfy the
IEC 60601-1.
►
Make sure that all configurations comply with
the system standard IEC 60601-1 chapter 16.
►
Do not bend the communications interface
cable, to prevent the cable from breaking and
the connection from failing.