Working With The Hilan Pneumatic Motor; System Set-Up; Connecting The Accessories; Connecting The Compressed-Air Supply - B.Braun Aesculap HiLAN GA519 Instrucciones De Manejo/Descripción Técnica

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 46
5.
Working with the HiLAN pneumatic
motor
5.1

System set-up

Injury to the operator and damage to the device
components can result from accidentally putting
the HiLAN pneumatic motor into operation!
WARNING
Use the safety stop to prevent any accidental
activation of the HiLAN pneumatic motor.
Only attach handpieces, tools, and the com-
pressed-air supply when the HiLAN pneumatic
motor is shut off.
Risk of infection and contamination! The HiLAN
pneumatic motor, its accessories, and the hand-
pieces are delivered in an unsterile condition.
WARNING
Sterilize the HiLAN pneumatic motor, its acces-
sories, and the handpieces before putting them
into operation.
5.1.1

Connecting the accessories

Combinations of accessories that are not mentioned in the present
instructions for use may only be employed if they are specifically intended
for the respective application, and if they do not compromise the perfor-
mance and safety characteristics of the products.
Please contact your B. Braun/Aesculap partner or Aesculap Technical
Service (address: see Technical Service) with any inquiries in this
respect.
Coupling the Hi-Line or Hi-Line-XS handpiece (e.g., GB519R,
GB520R–GB535R, or GB740R-GB758R)
Turn or slide safety stop 20 or slide 18 on the HiLAN pneumatic
motor 1 to position"0".
Turn threaded ring 24 in the direction of the arrow/rotation in
symbol
as far as it will go.
Attach the Hi-Line handpiece to the HiLAN pneumatic motor 1 so that
the nibs 21 on the HiLAN pneumatic motor 1 engage in the
notches 23 on handpiece 25. The handpiece can be mounted in 4 dif-
ferent positions.
Attach Hi-Line XS handpiece 26 to the HiLAN pneumatic motor 1 so
that the pin on the handpiece engages in one of the notches 22 on the
motor. The handpiece can be mounted in 15 different positions.
Turn threaded ring 24 in the direction of the arrow/rotation in
symbol
as far as it will go.
The handpiece is locked now.
Make certain that the threaded ring is fastened to its limit stop,
because otherwise it may come loose due to vibrations during opera-
tion.
Detaching the handpiece
Turn or slide safety stop 20 or slide 18 on the HiLAN pneumatic
motor 1 to position "0".
Turn threaded ring 24 in the direction of the arrow/rotation in
symbol
as far as it will go.
Remove handpiece 25 or 26 from the HiLAN pneumatic motor 1.
5.1.2

Connecting the compressed-air supply

Long tubes may cause a backup of exit air and thus a loss of performance
of the HiLAN pneumatic motor.
Keep tube lines as short as possible.
Make certain that the tube lines are not buckled or crushed.
Use a suitable extension tube if necessary, see Components required
for operation.
Note
The HiLAN pneumatic motors GA519/GA520R and GA529 including the
motor tube are sterile. Sterile separation takes place at the motor tube.
Connecting the HiLAN pneumatic motor at the foot control switch
Coupling:
Pull back unlocking sleeve (motor-side) 3 on motor tube 5 and connect
the motor tube with its fast-lock coupling (motor-side) 2 to the tube
connection 19 on the HiLAN pneumatic motor 1.
Release unlocking sleeve (motor-side) 3.
Insert connector (foot control-side) 6 of motor tube 5 into socket (foot
control-side) 8 on the foot control switch until it engages.
Uncoupling:
Pull back unlocking sleeve (motor-side) 3 on motor tube 5.
Remove fast-lock coupling (motor-side) 2 from the HiLAN pneumatic
motor 1.
Pull back unlocking sleeve (foot control-side) 7 on connector (foot
control-side) 6 of motor tube 5.
Pull connector (foot control-side) 6 out of socket (foot control-side) 8
on the foot control switch.
5

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Aesculap hilan ga520rAesculap hilan ga529

Tabla de contenido