Validated cleaning and disinfection procedure

Validated procedure
Specific requirements
Wipe disinfection for
electrical devices with-
out sterilization
LED light source
OP950
Manual cleaning with
immersion disinfection
Sterile adapter
OP941
Mechanical alkaline
cleaning and thermal
disinfection
Sterile adapter
OP941
8.6

Wipe disinfection for electrical devices without sterilization

Phase
Step
I
Wipe disinfection
RT:
Room temperature
Phase I
Remove any visible residues with a disposable disinfectant wipe.
Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
Observe the specified application time (1 min minimum).
Keep the optical cable interfaces clean.
Be sure that the optical cable interfaces are not scratched. Rub the
optical cable interfaces with a wad of cotton.
Suitable cleaning brush
20 ml disposable syringe
Drying phase: Use a lint-free cloth or medical compressed air
Place the instrument in a tray that is suitable for cleaning (avoiding
rinsing blind spots).
T
t
Conc.
[°C/°F]
[min]
[%]
≥1
RT
-
Water quality
Chemical
-
Meliseptol HBV wipes 50 % Propan-1-ol
Reference
Chapter Wipe disinfection for electrical devices
without sterilization
Chapter Manual cleaning/disinfection and sub-
section:
Chapter Manual cleaning with immersion
disinfection
Chapter Mechanical cleaning/disinfecting and
subchapter:
Chapter Mechanical alkaline cleaning and
thermal disinfecting