8.5.2
Validated cleaning and disinfection procedure
Validated procedure
Specific requirements
■
Wipe disinfection for
electrical devices with-
■
out sterilization
■
LED light source
OP950
■
Manual cleaning with
immersion disinfection
■
■
Sterile adapter
■
OP941
■
Mechanical alkaline
cleaning and thermal
disinfection
■
Sterile adapter
OP941
8.6
Wipe disinfection for electrical devices without sterilization
Phase
Step
I
Wipe disinfection
RT:
Room temperature
Phase I
►
Remove any visible residues with a disposable disinfectant wipe.
►
Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
►
Observe the specified application time (1 min minimum).
Keep the optical cable interfaces clean.
Be sure that the optical cable interfaces are not scratched. Rub the
optical cable interfaces with a wad of cotton.
Suitable cleaning brush
20 ml disposable syringe
Drying phase: Use a lint-free cloth or medical compressed air
Place the instrument in a tray that is suitable for cleaning (avoiding
rinsing blind spots).
T
t
Conc.
[°C/°F]
[min]
[%]
≥1
RT
-
Water quality
Chemical
-
Meliseptol HBV wipes 50 % Propan-1-ol
Reference
Chapter Wipe disinfection for electrical devices
without sterilization
Chapter Manual cleaning/disinfection and sub-
section:
■
Chapter Manual cleaning with immersion
disinfection
Chapter Mechanical cleaning/disinfecting and
subchapter:
■
Chapter Mechanical alkaline cleaning and
thermal disinfecting
11