instrumentation/parts of instrumentation be left in the patient during and following the
surgical procedure.
Inspect instruments for damage or wear prior to use. Failure to do so could result
in the damaging of implants or removal instruments.
* Note: If damage or wear is noted that may compromise the function of
the instrument, do not use the instrument and notify the appropriate
responsible person.
The following tables identify the torque ratings of the male hex and male hexalobe
drivers. The drivers should not be subjected to torque loads higher than the identified
ratings in order to ensure safe use.
3.2 mm (1/8˝)
Male Hexalobe Driver Torque Ratings
Male Hex Driver Torque Ratings
Size
1.5 mm
2.0 mm
2.5 mm
7/64˝
3.0 mm
3.5 mm
5/32˝
4.5 mm
3/16˝
5.0 mm
7/32˝
1/4˝
7.0 mm
Size
T08
T10
T15
T20
– 5 –
Torque Rating
(in.-lbs)
7
10
29
29
47
47
71
100
138
138
224
224
224
224
Torque Rating
(in.-lbs)
10
17
29
47