Given the aggressiveness of the active principles used in disinfectants, we recommend to use products containing the
following maximum quantities of:
● Ethanol at 96%. Concentration: maximum 30 g. for every 100 g. of disinfectant.
● Propanol. Concentration: maximum 20 g. for every 100 g. of disinfectant.
● Combination of ethanol and propanol. Concentration: combination of the two should be at most 40 g. for every
100 g. of disinfectant.
Cefla sc - Cefla Dental Group performed tests on the compatibility between the most common disinfectants and its own
plastic materials.
The test results indicated that the least aggressive agents are:
Incidin Spezial (Henkel Ecolab).
Omnizid (Omnident)
Plastisept ( ALPRO ) (not tuberculocide as not an alcohol-based disinfectant).
RelyOn Virkosept ( DuPont ).
Green and Clean SK ( Metasys ) (not tuberculocide as not an alcohol-based disinfectant)
The same tests showed that the above-mentioned products can therefore be used by adopting, however, the following
preATTENTIONs:
IMPORTANT INFORMATION !
Do not use products containing isopropyl alcohol (2-propanol,iso-propanol).
Do not use products that contain sodium hypochlorite (bleach).
Do not use cleaners that contain phenol.
Do not spray the chosen product directly on the equipment surface to disinfect.
Do not mix any of the above products nor mix them with any other products.
Any product should be used in accordance with its manufacturer's instructions.
Cleaning and disinfecting instructions
Clean and disinfect with single-use non-abrasive paper (avoid using recycled paper) or sterile gauze.
Sponge cloths or any other re-usable materials are not recommended.
IMPORTANT INFORMATION !
- In order to clean equipment connected to the power mains, shut off the devices and disconnect the power supply
38