NOTE:
When the On/Off switch is on and the Allen Advance Table is not
plugged into a power outlet, the Allen Advance Table operates on
battery power. Battery power is for short term use only and should be
used for emergency power situations only.
Battery charge level
Warning:
If the batteries are not charged before you
use the Allen Advance Table , the Allen
Advance Table may not function in an
emergency. Patient fall or injury could
occur.
When the Battery Charge Level indicator shows a low level of charge or if
you hear a "beep" from the Allen Advance Table when you try to use the
pendant controls, plug the Allen Advance Table into an applicable power
outlet to charge the batteries. To keep the batteries charged, plug the Allen
Advance Table into an outlet nightly and when you store the Allen Advance
Table.
NOTE:
A "beep" from the Allen Advance Table indicates that the battery
charge level is low or depleted.
NOTE:
When the Allen Advance Table is not plugged in: the Battery Charge
Level indicator will be on when the On/Off switch is on; the Battery
Charge Level indicator will be Off when the On/Off switch is Off.
Potential Equalization Conductor
There is a plug on the Allen Advance Table
for connecting an equipotential bonding
cable. This will equalise any potential
difference when the Allen Advance Table is
to be used with a patient who has intravenous,
cardiac, or other medical equipment.
Allen® Advance Table User Manual (D-720761 Ver D)
Instructions for Use
21