TILT THE SUPPORT TOP

NOTE:
For bariatric patients, we recommend that you install three supine
access tops (A-70404) to help support the patient during the tilt
procedure.
WARNING:
Before you adjust the Allen Advance Table or use the tilt
feature, make sure that the brakes are locked, the patient is
secured to the support top and is fully prepared for the
movement, and the Allen Advance Table is free of objects
that could interfere with the adjustment. Otherwise, patient
fall, injury, or equipment damage could occur.
Make sure that—
The brakes are locked (the locked indicator on
the Allen Advance Table display is green).
The patient is secured to the support top and is fully prepared for
the adjustment.
NOTE:
We recommend that you use the two large security straps to secure the
patient to the support top.
The Allen Advance Table is free of objects that could interfere
with adjustment.
WARNING:
Observe lines closely during articulations. To do so could
cause patient fall, injury, or equipment damage.
Press and hold the applicable tilt control until
the support top is at the correct angle. When you
release the control, the support top will lock at
that angle.
The tilt indicator on the display will show
how far the support top is tilted to the left or
right.
To return the support top to the flat position, press and hold the
opposite tilt control until the top is no longer tilted. The tilt indicator
on the display will be in the center of the tilt indicator area and the tilt
level indicator will be green.
Allen® Advance Table User Manual (D-720761 Ver D)