STEAM CLEAN
Do not use any steam cleaning device on the unit or put any Allen Advance
Table components in an autoclave or steam cleaning system. The unit or
mechanisms of the parts may be damaged due to excessive humidity.
DISINFECT
Caution:
Failure to use cleaning and disinfectant products in
accordance with the manufacturer's instructions could
cause equipment damage.
When there is visible soil and also between patient use, we recommend that
you disinfect the unit with an EPA registered (US only), tuberculocidal,
disinfectant such as Morning Mist Neutral Disinfectant.
Dilute the disinfectant as shown on the manufacturer's label.
TROUBLESHOOTING
WARNING:
Only Allen Medical-authorised persons should service the
Allen Advance Table. Service by unauthorized persons
could cause patient fall, injury, or equipment damage.
If the troubleshooting information below does not fix the problem, contact
Allen Medical Technical Support (see "SERVICE CALLS" on page 46).
THE ALLEN ADVANCE TABLE WILL NOT GO INTO
PENDANT CONTROL MODE FROM THE MANUAL
FLIP MODE
Make sure of these:
•
The brakes are locked.
•
The columns are at the same height.
•
The Allen Advance Table and the installed support top are level
(see "LEVEL THE ALLEN ADVANCE TABLE AND THE
SUPPORT TOP" on page 34).
•
The Allen Advance Table is free of objects that could interfere
with adjustment.
•
The flip handle is in the up and locked position
Allen® Advance Table User Manual (D-720761 Ver D)
Troubleshooting
45