Instructions for Use
i.
Push the flip handle up to the
locked position. You do not need
to press the red safety lever.
Determine which hole location on the H-bracket to install the support
top.
NOTE:
The hole locations are numbered on the side of the H-brackets.
WARNING:
Failure to use correct lifting methods could cause patient
fall, injury, or equipment damage.
With the help of another person, do as follows:
a.
Align the ends of the support top with the pre-determined hole
location. If you are to install a Flat Top without a pad installed,
make sure the side with the Velcro® tape is up (patient side).
NOTE:
You may need to adjust the length of the Allen Advance Table so that
the columns are at the correct distance apart to fit the support top (see
"ADJUST THE LENGTH OF THE ALLEN ADVANCE TABLE" on
page 23).
WARNING:
Never remove a cross pin from a support top while a patient
is on the support top. Patient fall or injury could occur.
b.
At each end of the support top,
insert a cross pin into the H-
bracket, through the support top
connector assembly, and through
the other end of the H-bracket.
WARNING:
Do not use the support top
without the cross pins installed
correctly through the H-brackets. To do so could cause
patient fall, injury, or equipment damage.
26
Allen® Advance Table User Manual (D-720761 Ver D)