Tracking Bracket
with Verza™ Guidance System
5. Remove Verza from guide storage column of gauge insert tree.
6. Using proper sterile technique, place Verza guide onto VerzaLink locating feature.
7. Rotate guide locking lever into locked position. Ensure guide is securely attached.
8. FIXED ANGLE APPROACH:
Using angle selector grip, slide angle indicator window to desired position and push angle selector lock
toward transducer to secure.
NOTE: Ensure angle selector lock is secure in desired position.
FREE ANGLE APPROACH:
Leave angle selector lock in unlocked position and instrument remains in-plane within constraints of guide
angle range.
9. Select appropriate size gauge insert and slide into guide body chute.
BOLT-ACTION QUICK-RELEASE
1. Rotate quick-release tab to open guide insert funnel.
REMOVING GUIDANCE SYSTEM
1. Rotate guide locking lever to unlocked position to remove guide from VerzaLink locating feature of bracket.
2. Remove and dispose of single-use components according to facility infection control policies.
3. Lift bracket latch to remove from transducer. Retain bracket for reprocessing.
25G, 22G, 21G,
20G, 18G, 17G,
16G, 15G, 14G,
12F, 10F, 9F, 8F, 7F