After proper cleaning and inspection, the PINPOINT camera may be sterilized
using the NOVADAQ Sterilization Tray (PC9018).
Maintain sterilization trays regularly and per manufacturer's instructions.
WARNING:
Follow sterilizer manufacturer's instructions.
WARNING:
WARNING:
Relevant, national/regional legal regulations must be observed.
CAUTION:
DO NOT use autoclave cycles to sterilize the PINPOINT camera.
To sterilize the PINPOINT camera using the NOVADAQ Sterilization Tray:
Insert three new filters provided in the sterilization accessories kit (PP9066):
two in the sterilization tray bottom and one in the lid (see Figure 57 and Figure
58).
The sterilization tray must be clean and dry prior to use.
WARNING:
Using the incorrect filters will result in the product not being properly
sterilized and may lead to the failure of the sterile barrier.
All three filters must be replaced with new filters prior to sterilization.
PINPOINT Endoscopic Fluorescence Imaging System Operator's Manual