ICG administration is to be performed via a central or peripheral venous line.
Use a three-way stopcock attached to an injection port on the infusion line.
Attach one prepared 3 ml syringe of ICG solution and one prepared 10 ml
syringe of saline solution.
Inject the prepared 2.5 mg/ml ICG solution into the line as a tight bolus.
Immediately switch access on the stopcock to the syringe containing saline
and briskly flush the ICG bolus through the line with 10 ml of sterile saline.
A fluorescence response should be visible in blood vessels visualized with the
PINPOINT System within 5 to 15 seconds after the injection.
Consumables should be disposed of in compliance with local, regional, and
national regulations.
Single-use or consumable components and accessories such as prepared or
partially used ICG should be disposed of in compliance with regulations for the
disposal of such items.