For a general troubleshooting guide see Table 12.
Table 12 Troubleshooting
Operating mode does not
change when illumination
button is pressed and
released.
Display option does not
change when display
button is pressed and
released.
"White balance failed" icon
is displayed after
attempting to white
balance PINPOINT
camera.
PINPOINT Endoscopic Fluorescence Imaging System Operator's Manual
PINPOINT is in
Standby
(illumination off)
mode.
PINPOINT is not in
Fluorescence
Imaging Mode
PINPOINT was
unable to adjust the
color response of
the camera.
If PINPOINT is in standby (illumination off) mode:
1.
Press the illumination button on the front
panel of the VPI.
2.
Press illumination button on the camera
to change operating mode.
If PINPOINT is in white light operating mode:
1.
Press the illumination button on the front
panel of the VPI.
2.
Press illumination button on the camera
to change operating mode.
If PINPOINT is in standby (illumination off) mode, see
previous troubleshooting item.
1.
Check that:
a. Camera and laparoscope are
connected to the VPI.
b. PINPOINT is in white light operating
mode.
c. The tip of the laparoscope is clean.
d. The tip of the laparoscope is being
held approximately 5 cm (2 inch) from
a matte white surface such as gauze
or white cloth when the white balance
switch is pressed.
2.
Retry the white balance.
If the white balance fails a second
time, contact a qualified NOVADAQ
service representative.
113