Note:
Ensure that the camera connector cap is secured over the connector during the
cleaning process.
Detailed manual cleaning instructions:
Prepare a fresh solution of enzymatic detergent
Wipe the entire surface of the device using a soft clean cloth dipped in the
detergent solution.
Immerse the device in the detergent solution, ensuring the solution contacts all
inner and outer surfaces.
Soak the device in the solution according to the manufacturer's
recommendations.
With the device immersed in the solution, thoroughly brush the exterior with a
soft-bristled brush, focusing on any mated or rough surfaces.
Actuate and brush any movable parts in all extreme positions.
Rinse each device with water
Once all detergent residue is removed, continue to rinse for 30 seconds.
Drain excess water from the device and dry it using a clean cloth or
pressurized air.
Visually inspect each device for cleanliness, paying close attention to hard-to-
reach areas.
1 Reverse Osmosis/Deionized water (RO/DI) at 30°C was used for validating cleaning efficacy.
2 Neodisher MediClean Forte will work at a ratio of 5-30 ml/L.
3 Neodisher MediClean Forte recommends a soaking time of 5 minutes.
PINPOINT Endoscopic Fluorescence Imaging System Operator's Manual
3
1
until all detergent residue is removed.
2
1
with lukewarm water
.
95