WARNING:
Do not use PINPOINT if any part of the system is damaged or does not
function properly. Failure to follow this warning may lead to injury.
PINPOINT contains no user serviceable parts. Do not attempt to open PINPOINT
components. Refer all servicing to a qualified NOVADAQ service representative.
As the system enters Standby mode, it will initialize the connected camera. If a
communication error is detected, the camera-failed error icon
and there will be no illumination (see Figure 62). If this error occurs:
Power off the VPI.
Disconnect and reconnect the PINPOINT camera, and then power on the VPI
again to clear the error.
If the camera fails a second time, turn the VPI power off and contact a qualified
NOVADAQ service representative.
Figure 62: Camera-failed icon
PINPOINT Endoscopic Fluorescence Imaging System Operator's Manual
Camera failed
will display
103