The following table uses the Spaulding classification to define the minimum level of
sterilization required for each PINPOINT component.
Table 8: Spaulding Classification and Minimum Reprocessing Requirements for PINPOINT
Laparoscopes
Nonsterile
Light Guide
Nonsterile
Cable
Camera
Nonsterile
VPI
Nonsterile
Cart
Nonsterile
Insufflator
Nonsterile
Monitor
Nonsterile
Recorder
Nonsterile
Printer
Nonsterile
Sterilization
Nonsterile
Trays
Introducer
Nonsterile
indicates the minimum required reprocessing level
1 The PINPOINT camera and light guide cable are classified as non-critical, but both must be sterilized because they enter the
sterile field and are subject to aseptic technique.
2 Clean the VPI or cart when visibly soiled and at regular intervals, as per a schedule established by the medical care facility or
provider.
3 Optional accessories vary based on regional and customer configuration.
4 Sterilization trays are classified as non-critical, but are sterilized during sterilization of PINPOINT components.
80
Components
Reusable
Critical
Reusable
Non-Critical
Reusable
Non-Critical
Reusable
Non-Critical
Reusable
Non-Critical
Reusable
Refer to Insufflator
Instructions
Refer to Manufacturer's
Reusable
Instructions
Refer to Manufacturer's
Reusable
Instructions
Refer to Manufacturer's
Reusable
Instructions
Reusable
Non-Critical
Single Use
DO NOT REPROCESS. Dispose after use.