Stryker PINPOINT PC9000 Manual Del Operador página 31

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 984
WARNING:
The following procedures must be performed using proper sterile technique
if they are being performed in preparation for surgery.
CAUTION:
The camera and laparoscope are precision medical instruments and must be
handled with great care.
Inspect the laparoscope for damage before and after use. If the laparoscope
is damaged, discontinue use and contact the manufacturer.
Do not subject to impact. Put the laparoscope down carefully.
Hold laparoscope only by the ocular funnel or body and not by the sheath.
Do not bend the sheath.
Store laparoscopes safely in a tray or similar container.
To connect the light guide cable to the laparoscope and VPI:
Attach the threaded light guide cable adapters to the laparoscope.
Connect the light guide cable to the threaded connector on the laparoscope.
Fully insert the other end of the light guide cable into the light guide cable port
of the VPI (see Figure 7).
Figure 7: Connecting the light guide cable to the VPI
PINPOINT Endoscopic Fluorescence Imaging System Operator's Manual
25

Publicidad

Tabla de contenido
loading

Tabla de contenido