Stryker PINPOINT PC9000 Manual Del Operador página 28

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 984
There are several NOVADAQ-certified recorders for use with the PINPOINT
System. Availability varies depending on region.
Note:
For the onscreen recorder controls and status to be visible to the operator during
normal operation of the PINPOINT System, the VPI SDI 1 video output must be
connected to the SDI-IN connector on the recorder. The SDI-OUT connector on the
recorder should be connected to the medical-grade video monitor. For more
information, see Figure 3.
For more information about setting up the recorder, please refer to Chapter 6.
WARNING:
All electro-medical devices connected to the PINPOINT system must be
certified medical-grade and all interconnected configurations shall comply
with the IEC/EN 60601-1 system standard. Failure to comply with this
standard may result in unsafe operation of the system and/or injury to the
patient or operator.
The VPI may be interconnected to other medical-grade video devices such as:
See Table 14 in Appendix C for specifications of the video output signals and
connectors and to ensure compatibility of the selected video devices.
Use the video-format switch on the rear panel of the VPI to select one of the
following video output formats:
Note:
Ensure that the video-format switch on the VPI is in the correct position for the
recorder that is being used.
22
Video recording, display or printing systems
Digital image-capture stations for hospital PAC systems
Stand-alone image management systems
HD-SDI 1080i 59.94
3G-SDI 1080p 59.94
4-0002428 Rev. J

Publicidad

Tabla de contenido
loading

Tabla de contenido