Issue
Cause
Loud running
Defective attachment
noise
drive/ball bearings
Attachment does
Attachment not compati-
not attach or
ble
detach.
Connection on attach-
ment deformed/defective
Coupling on drill
deformed/defective
Sleeve on drill hard to
move
Tool does not
Incompatible tool
attach or detach.
Tool connector
deformed/defective
Coupling on attachment
deformed/defective
Tool coupling dirty
For GB894R:
Kirschner wire chuck not
Kirschner wire
set correctly
does not insert
into the Kirschner
wire chuck
For GB894R:
Kirschner wire chuck not
Kirschner wire
set correctly
not turning
Open the clamping lever
Clamping lever not pulled
back with sufficient force
Tool not moving
Attachment not fully
attached to the drill
Tool not fully attached to
the attachment
Attachment defective
Drill defective
The safety catch of the
drill is in the OFF position
For GB896R/GB897R:
lever pulled
For
Lever mechanism defec-
GB896R/GB897R:
tive
changing
between
mechanical and
Ratcheting mechanism
manual opera-
defective
tion is not possi-
ble.
Oil spray adapter
Oil spray adapter not
not attachable
compatible
12
Detection
Loud, perceptible noise during operation
Attachment doesn't engage
Attachment is difficult to couple or can-
not be coupled
Attachment is difficult to couple or can-
not be coupled
Attachment doesn't engage
Tool does not click into place
Tool is difficult to couple or cannot be
coupled
Tool is difficult to couple or cannot be
coupled
Tool is difficult to couple or cannot be
coupled
Diameter of the Kirschner wire is not
consistent with the scale setting
Diameter of the Kirschner wire is not
consistent with the scale setting
Open the clamping lever
Kirschner wire not turning
Attachment can be pulled out of the drill Correctly attach the attachment and perform a function
Tool can be pulled out of the tool cou-
pling
Drill turning, but attachment is not turn-
ing.
Drill not turning
Safety catch is in OFF position
For GB896R/GB897R: lever pulled
Lever is difficult to move or cannot move.
Lever is bent.
Lever does not move itself back to the
starting position.
Lever is difficult to move or cannot move.
No ratcheting function in manual opera-
tion.
Oil spray adapter not attachable
Remedy
Follow operating instructions (preparation, care).
Preventative: lubricate attachment before every steriliza-
tion.
Have the manufacturer repair the attachment.
Use the compatible attachment for drill.
Have the manufacturer repair the attachment.
Have the manufacturer repair the drill.
Turn and hold the drill sleeve, then attach the attachment.
Turn the sleeve back if necessary.
Preventative: lubricate the sleeve on the drill before each
sterilization.
Use the tool compatible with the attachment.
Use a new tool.
Have the manufacturer repair the attachment.
Clean tool or use a new tool.
Clean the attachment.
Set the adjustment sleeve to the right Kirschner wire diam-
eter.
Set the adjustment sleeve to the right Kirschner wire diam-
eter.
Close the clamping lever.
Pull back the clamping lever forcefully and hold it.
test.
Correctly attach the tool and perform a function test.
Have the manufacturer repair the attachment.
Have the manufacturer repair the drill.
Turn the safety catch into the ON position.
For GB896R/GB897R: do not pull lever for mechanical oper-
ation.
Clean and oil attachment.
Have the manufacturer repair the attachment.
Alternatively, use a screwdriver handle.
Clean and oil attachment.
Have the manufacturer repair the attachment.
Alternatively, use a screwdriver handle.
Use the compatible oil spray adapter for GA344.