7 Mechanical conversion
7.14 Installing the 3D module
K
D
C
G
F
38
Converting instructions for the "ORTHOPHOS XG 5 to XG 3D upgrade kit" ORTHOPHOS XG 5
4. Screw the cover tight to the rotation unit with the two screws (K)
(included with the installation material).
Re-tighten screw B.
K
5. Hook the 3D module (C) onto the rotation unit (D) and secure it
immediately with the 3 screws (E).
E
6. Attach the sheet metal bracket (F) (incl. cables and strain reliefs) to
the 3D module using the 3 screws (G).
Sirona Dental Systems GmbH
63 52 970 D3352