8 Updating the unit software and configuring the unit
Adding a unit as an XG 3D X-ray
component
Saving data from board DX61 to board
DX11
Saving data to board DX88
50
Converting instructions for the "ORTHOPHOS XG 5 to XG 3D upgrade kit" ORTHOPHOS XG 5
1. Call the SIDEXIS Manager, delete the XG X-ray component and add
the unit as an XG 3D X-ray component.
2.
NOTICE! Now the unit is identified as ORTHOPHOS XG 3D.
Repeat the software update to the current unit software version (in
update mode
ORTHOPHOS XG 3D Service Manual, "General operating
sequences → Software update".
➢ Call service routine S017.17 and transfer the data from board DX61
to board DX11 (see ORTHOPHOS XG 3D Service Manual, "Service
→ S017: Test step 17", Case 1).
IMPORTANT: The data direction shown in selection field 3 must be
DX61 → DX11.
1. Call service routine S009.8 (saving/restoring DX88 data) and execute
the routine (see ORTHOPHOS XG 3D Service Manual, "Service →
S009: Test step 8").
2. Switch the unit off.
Wait approx. 1 minute. Then switch the unit on again.
The unit software has now been updated and the unit is configured.
Following a unit restart, only error message E1 11 20 (unit is not
calibrated) should appear.
"Automatic"
). Proceed as described in the
Sirona Dental Systems GmbH
63 52 970 D3352
D3352.092.08.05.02 07.2012