Converting instructions for the "ORTHOPHOS XG 5 to XG 3D upgrade kit" ORTHOPHOS XG 5
Confirming the unit serial number
Defining the unit version
Entering the activation key code
Selecting a language and language set
63 52 970 D3352
D3352.092.08.05.02 07.2012
Once the key combination has been entered correctly, service
routine S017.1 (select/confirm unit class) is started automatically.
The memory key (A) lights up.
4. Confirm the unit class "01":
To do so, touch the memory key (A) (R key on the touchscreen lights
up) and then the R key on the Touchscreen (B).
5. Exit the service routine with the double-arrow key (C).
6. Switch the unit off.
Wait approx. 1 minute. Then switch the unit on again.
Error message E61505 (undefined unit serial number) is
displayed.
1. Acknowledge the error message using the R key on the control panel.
2. Call the Service menu
(see ORTHOPHOS XG 3D Service Manual).
3. Call service routine S008.3 (see ORTHOPHOS XG 3D Service
Manual, "Service → S008: Test step 3").
4. Check the serial number
IMPORTANT: The unit serial number is located on the nameplate of
the unit.
NOTICE! If the serial number is incorrect, cancel the update and
contact the Sirona Customer Service Center!
5. Confirm the serial number using service routine S008.3 (see
ORTHOPHOS XG 3D Service Manual, "Service → Basic operating
sequences in the Service menu").
6. Switch the unit off.
Wait approx. 1 minute. Then switch the unit on again.
1. Call service routine S017 and set the unit version using service
routines S017.2 (see ORTHOPHOS XG 3D Service Manual, "Service
→ S017: Test step 2").
2. Switch the unit off.
Wait approx. 1 minute. Then switch the unit on again.
"No Key"
The message
1. Enter the activation key code using the Easypad (see ORTHOPHOS
XG 3D Operating Instructions, "Settings and repairs → Activate
function → Activation via Easypad").
2. Switch the unit off.
Wait approx. 1 minute. Then switch the unit on again.
"No Key"
The message
➢ Call service routine S017 and set the language and the language set
using service routines S017.4 and S017.5 (see ORTHOPHOS XG 3D
Service Manual, "Service → S017: Test step 4 and 5").
8 Updating the unit software and configuring the unit
may appear on the Easypad.
must no longer appear.
49