Converting instructions for the "ORTHOPHOS XG 5 to XG 3D upgrade kit" ORTHOPHOS XG 5
63 52 970 D3352
D3352.092.08.05.02 07.2012
9
Adjusting/calibrating the unit
Adjusting/calibrating the unit
1. In SIDEXIS XG, call the
"Utilities"
"Constancy test..."
/
"Service exposure "
calibration"
/ Password request (see also Service Manual, chapter
"Adjusting/calibrating the unit")
IMPORTANT
Before you begin adjusting or calibrating the unit, you must reset the 3D
adjustments.
2. In the structure tree, under
adjustment"
element.
"Reset values"
3. Click on the
IMPORTANT! After resetting the values, the parameter C1 must
have a value of "88".
4. Now perform a complete unit adjustment/calibration, as described in
the ORTHOPHOS XG 3D Service Manual, "Unit adjustment/
calibration" chapter.
9 Adjusting/calibrating the unit
"Adjustment/calibration"
"3D"
"Select X-ray device"
/
/ (
"Select X-ray component"
/ (
"3D Reset adjustment"
button.
menu:
) /
"Adjustment/
) /
"Reset
, click on the
51