7 Mechanical conversion
7.7 Removing the Pan sensor holder
28
Converting instructions for the "ORTHOPHOS XG 5 to XG 3D upgrade kit" ORTHOPHOS XG 5
7.7
Removing the Pan sensor holder
Removing the Pan sensor holder
Removing the Pan sensor holder: XG5/XG 3D Ready
C
D
E
1. CAUTION! Cable L13!
Loosen the three screws (C) and remove the sensor holder.
2. Loosen the two screws (D) and remove the "small sensor holder"
cover.
3. Loosen the three screws (E) and remove the connection socket.
4. Remove cable L13 from the shield terminal (F) at the rear of the
connection socket.
5. Loosen the connector screw connection on the front side of the
connection socket and thread the connector to the rear through the
connection socket and the sensor holder.
NOTE: The removed sensor holder is no longer required for reassembly.
Dispose of the sensor holder in accordance with the regulations
applicable in your country [ → 58].
Sirona Dental Systems GmbH
63 52 970 D3352
D3352.092.08.05.02 07.2012