8 Updating the unit software and configuring the unit
Perform a software update
Confirming the unit class
48
Converting instructions for the "ORTHOPHOS XG 5 to XG 3D upgrade kit" ORTHOPHOS XG 5
8
Updating the unit software and configuring
the unit
Updating the unit software and configuring the unit
IMPORTANT: The current update file can be found on the ORTHOPHOS
XG3D Firmware CD. This is included in the XG 3D conversion set (IT
package) and also the country-specific set. The contents of the CD can
be downloaded from the Dealer domain of the Sirona Internet home page
(under Product Info/X-ray Systems): www.sirona.com
Update: XG5 to XG 3D
✔
The functionality for use of the Easypad must be activated [ → 17].
✔
The IT package from the XG3D upgrade kit must be installed.
✔
The unit is mechanically converted to ORTHOPHOS XG 3D.
"Automatic"
1. In
to the latest unit software version, as described in the ORTHOPHOS
Plus
XG5/XG
NOTICE! At this point, do not acknowledge any error messages on
the unit or the PC (e.g. "Timeout suspected").
2. Switch the unit off.
Wait approx. 1 minute. Then switch the unit on again.
Error message E6 11 07 (undefined system class) is displayed.
1. Acknowledge the error message with the R key .
The access level for the service menu (level 4) is automatically
started.
2. Press and hold down the service key (D) until the patient symbol keys
light up (approx. 2 s).
3. Then press the patient symbol keys in the sequence F – H – E within
the next 4 s.
update mode, perform an automatic software update
ready
and XG 3D
Service Manual.