Disposal of removed parts
16
Converting instructions for the "ORTHOPHOS XG 5 to XG 3D upgrade kit" ORTHOPHOS XG 5
1. Adjust the FH (Pan) double laser light localizer (see ORTHOPHOS
XG 3D Service Manual, "Repairs → Laser light localizers" chapter).
2. Adjust the MS (Pan) laser localizer (see ORTHOPHOS XG 3D
Service Manual, "Repairs → Laser light localizers" chapter).
➢
Perform a complete unit adjustment or calibration [ → 51].
➢ Complete the unit conversion as described in the chapter entitled
"Final work [ → 52]" .
➢ Dispose of parts which are no longer required as described in the
chapter entitled "Disposal/returns [ → 58]" .