Converting instructions for the "ORTHOPHOS XG 5 to XG 3D upgrade kit" ORTHOPHOS XG 5
Activation key code
Activating the functionality for the
Easypad
Installing the IT package
Mechanical conversion
Updating the unit software and
configuring the unit
63 52 970 D3352
D3352.092.08.05.02 07.2012
5
Procedure for conversion
Procedure for conversion
IMPORTANT
The unit may only be converted once a function check [ → 12] has been
performed and no deficiencies have been found on the unit.
Tip: Sirona recommends calling [ → 53] the "Extended details" before
starting the conversion in order to read out the current parameters from
the unit and storing these as an XML file under the network name of the
unit in the PDATA/.../P2K_Config directory for documentation purposes.
Activation key: XG5/XGPlus
➢ Have the activation key code ready for the conversion.
The activation key code can be requested from the Sirona Customer
Service Center (CSC).
S008.30: XG5
➢ Activate the functionality for later use of the Easypad [ → 17].
➢ Install the IT package (see GALILEOS / ORTHOPHOS XG 3D
Software Installation technical documentation (REF 61 42 389)). This
can be found on the documentation CD in the IT package.
1. Remove the Pan sensor and move the unit to a comfortable working
Mechanical conversion: XG5
position [ → 20].
2. Remove the cover parts [ → 20].
3. Remove the head support and outer ring [ → 23].
4. Replace the ring motor [ → 24].
5. Remove the old diaphragm unit [ → 25].
6. Remove the old X-ray tube assembly [ → 26].
7. Remove the Pan sensor holder [ → 28].
8. Install the ferrite cores [ → 29].
9. Install the connection socket for cable L13 in the housing [ → 30].
If not yet laid:
10.
Lay light barrier cable L21 and motor cable L20 on the ring in the
brackets of the sliding plate [ → 31].
If not yet laid:
11.
Lay [ → 32] fiber-optic cable L15.
12. Attach cable ties to secure the spiral springs [ → 35].
13. Install and connect the wires for the rotation unit [ → 36].
14. Install and connect the wires for the 3D module.
15. Insert the 2D sensor.
16. Install and connect the wires for the new XG 3D X-ray tube
assembly [ → 42].
17. Install and connect the wires for the diaphragm unit.
18. Replace the user interface [ → 46].
19. Install the outer ring and head support [ → 47].
20. Attach the cover parts [ → 47].
Update the unit software and configure the unit [ → 48].