7 Mechanical conversion
7.6 Removing the old X-ray tube assemblyConverting instructions for the "ORTHOPHOS XG 5 to XG 3D upgrade kit" ORTHOPHOS XG
26
7.6
Removing the old X-ray tube assembly
Removing the old X-ray tube assembly
1. Loosen the four screws (A) and remove cover plate (B) incl. the cable
shielding (L3). CAUTION! Also pull cable L3 off connector X3 and the
ground cable off connector X304 on board DX6.
Tip: The ferrite core and cable shielding can remain on the cover
plate.
2. Detach cables L5 and L6 from the rubber grommets and pull the
cables out of sockets J6 (L5) and J2-J3 (L6) on board DX6.
3. Loosen the two rear screws (C) on the tube assembly.
Sirona Dental Systems GmbH
63 52 970 D3352
D3352.092.08.05.02 07.2012