8 Care and cleaning

8
SIROEndo Pocket unit/
Other components
32
Care and cleaning
The unit and its accessories must be cleaned and disinfected and the muco-
sal electrode and the file clamp of the apex locator must be sterilized prior to
initial operation and before treating each patient.
Unplug the power supply unit from the power outlet prior to cleaning and/or

disinfection.

Do not apply any liquids to the unit or spray them directly onto it.
Never immerse the unit or its accessories in liquids.
The packaging of the unit or its accessories must not be used as sterilization
packages.
8.1

Cleaning

The unit and the control panel as well as the micromotor, micromotor
holder, micromotor cable and accessories can be wiped off with a cloth
moistened lightly with water.
To remove stubborn dirt and grime, clean the unit and the control panel
as well as the micromotor and hose with a piece of cloth moistened slight-
ly with a neutral cleaning agent and then wipe it dry with a soft cleaning
rag.
Automatic disinfection is not permissible.
Do not clean in an ultrasonic cleaning system.
The micromotor, foot switch and all cables can be wiped off with a cloth
moistened lightly with water or a cleaning agent.
Never use any solvents or aggressive cleaning agents to clean the micro-
motor, the foot switch or the cables.
Remove dirt, grime and disinfectant residue regularly using mild, com-
mercially available cleaning agents.
Do not allow liquids to penetrate into the slots of the unit!
Immediately wipe off medications that come into contact with the surface.
8.2
Disinfection
The unit and the control panel as well as the micromotor, micromotor
holder, micromotor cable and accessories can be wiped off with a cloth
moistened lightly with a disinfectant.
Never use sprays near the connectors or connecting pieces.
You must choose only disinfectants that comply with the valid require-
ments of the respective national regulatory body or whose bactericidal,
fungicidal and virucidal properties have verifiably been tested and ap-
proved accordingly. You can use for instance:
MinutenSpray classic from Alpro, MinutenWipes from Alpro.
In the USA and Canada:
®
CAVICIDE
or CAVIWIPES
Do not use: Substances containing phenol, peracetic acid, peroxide or any
other oxygen-splitting agents, sodium hypochlorite or iodine-splitting agents.
Operating Instructions SIROEndo Pocket
D 3526.201.01.05.02 07.2012