9 Maintenance
Apex locator accessories
9
Micromotor
34
The mucosal electrode and file clamp can be sterilized for 15 minutes at a
steamtemperature of 134 °C (273 °F).
Steam sterilizers that fulfill the requirements of EN 13060, Class B and are
suitable for the sterilization of micromotors are approved for sterilization.
CAUTION: The apex locator cable for connection to the
SIROEndo Pocket cannot be sterilized.
WARNING!
A temperature of 134 °C (273 °F) must not be exceeded during steriliza-
tion.
Do not immerse the components in any solutions.
Do not use ultrasonic cleaning systems!
Proper maintenance and periodic checks of the function and tempera-
ture of the sterilization system are important.
Maintenance
The micromotor is maintenance-free:
Never oil the micromotor!
Do not insert any objects in the micromotor!