Manual cleaning with immersion disinfection
Phase
Step
I
Cleaning
II
Rinse 2x
III
1)
Disinfection
IV
1)
Rinse 2x
V
Final Rinse
VI
Drying
FD-W
Fully desalinated water (demineralized, low microbiological contamination, max. 10 germs/ml and low in endotoxins, max. 0.25 Endo-
toxin units/ml)
1)
Phase does not apply to US market
Damage to the product due to improper handling!
Scratch-sensitive surface.
CAUTION
Handle the device with appropriate care.
Do not use metal brushes, metal objects or abra-
sive cleaners.
Phase I: Cleaning
Prepare the cleaning solution according to the manufacturer's instruc-
tions.
Fully immerse all parts of the fully disassembled device that were used
in the cleaning solution bath for two (2) to five (5) minutes. All acces-
sible surfaces must remain immersed in the cleaning solution bath
throughout the entire cleaning time.
After soaking in the cleaning solution, clean all external surfaces-
while the items are fully emerged in the cleaning solution – using a
clean, lint-free, soft, non-sterile cloth or a soft bristled brush until all
visible soiling is removed:
– Brush for at least one (1) minute or until no more residues can be
removed.
– During cleaning, move non fixed components 3 times in each direc-
tion and as far as possible.
Thoroughly rinse all surfaces of products with hidden crevices, lumens
with working channels, or complex geometry five (5) times. Use a dis-
posable 50 ml syringe.
Phase II: Rinse
Fully immerse all parts in a tap water bath (<45 °C/113 °F) and thor-
oughly rinse all accessible surfaces twice for a minimum of one (1)
minute each.
Move non fixed components 3 times in each direction and as far as
possible.
T [°C/°F]
t [min]
as per manufac-
2-5
turer's instruction
<45/113
2x ≥1
as per manufac-
12
turer's instruction
<45/113
2x ≥1
<45/113
≥1
Water quality
Chemicals
Tap water
Enzymatic cleaning agent (Cidezyme/Enzol)
Tap water
as per manufac-
0.55 % Orthophtaldehyde solution (Cidex OPA)
turer's instruction
Tap water
FD-W
Thoroughly rinse all hidden crevices, lumens or complex geometry at
least three times:
– Use a disposable 50 ml syringe.
– Use fresh water for each rinse.
– Allow water to drip off for a sufficient length of time.
Phase III: Disinfection
Immerse all parts completely in disinfectant solution for at least
12 minutes. All accessible surfaces must remain immersed in the dis-
infectant solution bath throughout the entire disinfecting time.
Remove all adherent air bubbles from the component surfaces
Thoroughly rinse all surfaces of products with hidden crevices, lumens
with working channels, or complex geometry five (5) times. Use a dis-
posable 50 ml syringe.
Move non fixed components 3 times in each direction and as far as
possible.
Phase IV: Rinse
Fully immerse all parts in a tap water bath (<45 °C/113 °F) and thor-
oughly rinse all accessible surfaces twice for a minimum of one minute
each.
Move non fixed components 3 times in each direction and as far as
possible.
Thoroughly rinse all hidden crevices, lumens or complex geometry at
least three times:
– Use a disposable 50ml syringe.
– Use fresh water for each rinse.
– Allow water to drip off for a sufficient length of time.