Immunity Test
IEC 60601-1-2 Test Level
Electrostatic
+ 6 kV contact
discharge (ESD)
+ 8 kV air
IEC 61000-4-2
Electrical fast
+ 2 kV for power supply lines
transient/burst
+1 kV for input/output lines
IEC 61000-4-4
Surge
+ 1 kV differential mode + 2
IEC 61000-4-5
kV common mode
Voltage dips, short
< 5% U
interruptions, and
voltage variations
40% U
on power supply
input lines
70% U
IEC 61000-4-11
< 5% U
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
Note: U
Electromagnetic Immunity - General
(> 95% dip in U
)
T
T
for 0.5 cycle
(60% dip in U
)
T
T
for 5 cycles
(30% dip in U
)
T
T
for 25 cycles
(> 95% dip in U
)
T
T
for 5 sec.
is the AC mains voltage prior to application of the test level.
T
Electromagnetic Compatibility (M3813B Scale)
Compliance
Electromagnetic
Level
Environment Guidance
+ 6 kV contact
Floors should be wood,
+ 8 kV air
concrete, or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be at
least 30%.
Not Applicable
Not Applicable
Not Applicable
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Specifications and Regulatory
5-15