Descargar Imprimir esta página

Philips Patient Telemonitoring Set Instrucciones De Uso página 68

Publicidad

Idiomas disponibles

Idiomas disponibles

Replacing Batteries
The display will then show
flash slowly.
Step 7.
When the display is properly connected to the Scale:
Note
Important: Do not step on the Scale while doing this step.
If the Step Off light does not stop flashing, check the cord connections
by unplugging the cord and replugging it in until it snaps into place.
Also check that the batteries are inserted the right way.
Step 8.
Caution
Do not lift the Steady Scale.
If the Steady Scale must be moved, ask for assistance or call your
healthcare provider for help.
Step 9.
Step 10.
Note
If the Scale does not seem to be operating properly or the Radio Test
fails, contact your healthcare provider.
3-8
Maintenance
Replace the display as follows:
– Place the connection cord through the display cutout.
– Hold the scale display above the top of the column with the
display facing you.
– Plug the cord into the adapter inside the top of the column.
– Center the display over the top of the column so that the plastic
fastener strips on the display and the column match.
– Press down on the display. The back of the display should rest
against the rim on the top of the column.
– the Step Off light stops flashing.
– the Scale will say "zero pounds", display 0.0, and then turn off.
Return the Scale to its original spot.
Perform the Radio Test described in the Troubleshooting
section.
Step on the Scale and verify that it is working properly.
briefly, and the Step Off light will

Publicidad

Capítulos

loading